Characteristics:
Saturation rate 4Three pulse rate
Low injection mode
In your simcube box
Run the aa battery。
Have sleep mode。
Correct reading
Compatible sensor testing
Nell / mas settings:
All nellcor
Masimo
Datascope
Agreements
Spacelabs
Hp / philips
Hp / philips
Nihon kohden
Colin
Ge / marquette
Welch allyn
Hp / om settings:
Hp / philips
Ohmeda
Novametrix
Mode:
85%: saturation of simulated patients is 85%, heart rate 80bpm, injection index approximately 2. 0
95 per cent: saturation of simulated patients is 95 per cent, heart rate 40bpm, injection index approximately 2. 0 note: for many blood oxygen, the heart rate is within the limits of performance specifications and some may be in trouble。
98 per cent: saturation of simulated patients is 98 per cent, heart rate is 80 bpm, injection index is about 2. 0
140 bpm: 98% saturation of simulated patients with a heart rate of 140 bpm and an injection index of approximately 2. 0
Low injection: simulated saturation of 99 per cent, heart rate of 80bpm and injection index of approximately 0. 2 patients. Note that this model simulates challenging low injection levels and does not apply to all monitors or sensors。
Spo2 sensor attention:
The long-lasting spo2 sensor has a public transport specification of +/- 2 per cent, the one-time spo2 sensor has a public transport specification of +/- 2 per cent or +/- 3 per cent, and the additional public transport specification of +/- 1 per cent for most haemobioxymeters. This indicates that the theoretical overall error was up to 4 per cent, but in most cases the haematological saturation reading should be within 2 per cent of the value indicated on oxsim。
Ox-1 oxsim optical spo2 simulator, ox-1 oxsim pulse blood oxygen simulator operational description:
Sensor led light: oxsim has two leds marked respectively as "ir" and "red" near the bottom of its display screen. When a large amount of infrared light is detected on the input side of the finger, the led marked “ir” is lighted, and the led marked “red” is lighted when a large number of red light is detected on the input side of the finger. Both leds should be stable and bright (not blinking) when the oxsim is connected to the normal work of the saturation meters and sensors. These leds are usually only checked to detect sensor failure。Resuscitation mode: when no sensor is connected to oxsim in 10 minutes, it will turn off itself to save electricity. To reopen it, you can close its power and then open it, or press one of the mode or type buttons. Oxsim will not remember your previous mode or type settings, so when you reopen them, please reconfigure them。
Battery power indicator: oxsim will perform active spo2 simulations on single aa batteries for approximately 8 to 12 hours, depending mainly on the brightness of the light from the serooxymeter. The battery indicator will become green when the battery is in good condition (the load voltage is in the 1. 6v to 1. 1v range ... Approximately 80% of the total running time). The battery indicator will be orange when the battery power is critical (the battery voltage loaded is in the range of 1. 1v to 1v ... Approximately 15% of the total running time). When battery power is exhausted, the battery signal will be red. When the battery signal light is red, oxsim may be activated and appear to be functional, but may be reset with the spo2 sensor
Ox-1 oxsim optical spo2 simulator, ox-1 oxsim pulse blood oxygen emulator, using oxsim optical spo2 pulse oxygen simulator, tested the entire spo2 system, including sensors and extended cables. In the pocket of the shirt, you can easily operate with only two buttons for a section aa battery. Automatic detection of sensor types eliminates the need to select the manufacturer of the blood detector. An optional serial communication port allows users to expand saturation, pulse rate and emulation settings, which are well suited to the need for extensive simulation of actual applications。
Oxsim optically saturated blood pulser tester characteristics:
Spo2 simulation:
Mode 1:
Saturation: 85 per cent
Heart rate: 80 bpm
Injection index: around 2. 0
Mode 2:
Saturation: 95 per cent
Heart rate: 40 bpm
Injection index: around 2. 0
Mode 3:
Saturation: 98 per cent
Heart rate: 80 bpm
Injection index: around 2. 0
Mode 4:
Saturation: 98 per cent
Heart rate: 140 bpm
Injection index: around 2. 0
Mode 5:
Saturation: 99 per cent
Heart rate: 80 bpm
Injection index: around 0. 2
Dimensions:
2-1 / 2 inches x 2-1 / 2 inches x 1-1 / 4 inches (6. 4 x 6. 4 x 3. 2 cm), excluding finger parts
Weight:
Just five ounces. (0. 14 kg, excluding batteries)
Operation simple:
Just two buttons
Facilitate:
No need to purchase or carry adapter cables
Economy:
The "finger" is fixed and permanent. Spare parts not required to be carried or replaced each year
Rustling:
Polycarbonate shell. Down from a height of 3 feet (1m) to a hard surface with 65 tests, no faults
Power:
6vdc adapter or aa battery. Battery boost – runs oxsim and simcube nibp simulator。
Battery lifetime:
At least eight hours. 60 days in sleep mode
Pledge:
Four years limited warranty
Packaging:
(1) oxsim optical o2 saturation pulser tester
(1) power (6vdc)
(1) aa batteries
(1) calibration certificate
(1) boxes
Optional function:
Serial communication ports provide user control over oxsim for extended simulation settings。
Saturation and scope
10-100 per cent
Pulse range
10-250 bpm
Injection index range
0. 1 – 2. 0 pi