Dp-600 controls by touch screen or pc:
- when testing defibrillators to perform single-phase, two-phase and pulse-phase energy measurements, they can be tested in synchronous and helipad mode and measured in 50 internal loads
- test sheets, double- and triple-circuit pacemakers can measure voltage, range, time, frequency, pulse...
-ecg simulations can provide pulse output for defibrillators and ecg equipment
Spl dp-600 defibrillator, dp-600 bottling analyser characteristics:
- Display: touch 4. 3 "tft-display
- Graphical display of discharge process when using pc
- Interface: 1 x usb 3. 0
- Measurement of recharge time
- Powered by municipal or built-in batteries
- 8 language menus (german, english, french, polish, spanish, italian, portuguese, turkish)
Other functions:
- Power voltage: 18 v al. Internal batteries
- Working temperature: +5 to +40°c
- Storage temperature: -10 to +50°c
- Dimensions: 235 x 90 x 330 mm
- Weight: approximately 2. 5 kg
The standard configuration of dp-600 includes:
- Calibration certificate
- User manual on cd
- Network
- Usb cable
- 10 sta8 ecg adapters check
- 2 propeller sensor components with an integrated 4 mm plugin for defi
- 6 4mm pace outlets
- 10 4 mm ecg outlets
Spl dp-600 defibrillator, technical parameters for dp-600 bottler analyser:
| Measurement | Scope | Precision |
| "defi." | Step, step, step, step | |
| Barrier measurement | 50 times | ± 1% |
| Dynamic | 1 to 1000 j | ±1 j (1% of measured) |
| Pulse | 0 to 403 v / scope 1 / | ±1 v (1% of measured) |
| 0 to 4440 v / scope 2 / | ±1 v (1% of measured) | |
| Pulse width | 0 to 20 ms | ± 0. 1 ms (2% of measured) |
| Pulse delay | 0 to 100 ms | ± 0. 1 ms (2% of measured) |
| Sensitivity | One | |
| Step | Through the chest, inside the heart | |
| Barrier measurement | 50 to 1600 (50 ek steps) | |
| Pulse | 0. 1 to 277. 5 v | ± 0. 1 v (1% of the measured value) |
| Pulse width | 0. 1 to 250 ms | ±1 ms (5% of measured) |
| Frequency | 30 to 800 bpm | ±1 bmp (1% of the measured value) |
| Av delay | 10 to 400 ms | |
| Frequency required | 55 to 65 bpm | |
| Time out | 50 to 400 milliseconds | |
| Sensitivity | 0. 5 to 25 mv | |
| "the e. G." | 12 channels | |
| Pulse course | Nostrils, square nostrils, triangles, quadrants, flying men | |
| Room tremors (vf), room hysteria (vt) | ||
| Iso, line, qrs |