Since its launch, rotroNic hygrogen defines the criteria for the calibration of portable humidity and temperature. Hundreds of users around the globe have realized that such tools for rapid generation of stable temperature and humidity conditions can save significant time for calibrating all types of humidity instruments of all manufacturers。
Hygrogen2 hg2-s
Hygrogen2 has been able to calibrate instruments throughout its work, thus further meeting stringent quality and compliance requirements, and is considered to be the leading instrument in the same product throughout the pharmaceutical industry and is favoured by laboratories around the world in the iso 17025 wet calibration. Portable and robust for durable use, which can be configured at the installation location for complete system identification。
Hygrogen2 hg2-xl
The stability of hygrogen2-s was replicated as a rock in hygrogen2-xl, with a minimum heat gradient and a rapid change in setting points, with 10 times the processing room capacity of the product。
Features
- Generates a steady humidity of 2,99% rh and temperature of 560° condition (<0. 0. 1% rh, <±0. 01°c)
- Humidity and temperature change in less than 5 minutes
- Time-saving multiprompt calibration
- User programmable calibration sequence (up to 200 steps)
- Automatic calibration (optional)
- Integrated rotro matching fda 21 cfr part 11Nic hw4 software
Apply
- On-site calibration solutions for all users of humidity and temperature measurement devices。
| Hg2-s | Hg2-xl | |
| Type | Standard size humidity and temperature calibration | Xl size humidity and temperature calibration |
| Climate scope | 0... 60°c, extension 5... 60°c 5... 95% rh, spread to 2... 99% rh |
|
| Terms of reference | 10-35°c, altitude < 2000 m | |
| A probe | Six detectives through the door | Up to 19 probes through the door, and the interior shelf |
| Workload | 1. 5 litres | 17 litres |
| Typical calibration uncertainty | 23 °c § 1. 5% rh(k = 2), ± 0. 15 °c (k = 2) 15 ... 50 °c | |
| Humidity change 5... 95% rh | < 5 minutes | < 15 minutes |
| Temperature change 23... 50°c | < 5 minutes | < 15 minutes |
| Temperature change 23... 0°c | < 25 minutes | < 35 minutes |
| Programmer function | 20 user programs with a maximum of 200 steps per program | |
| External sampling rings | The temperature control exit and entrance, 6 mm | |
| Optional enhancements | Temperature/moisture range extension, autocal, external mbw / rhs Reference integration, remote screen sharing, remote api |
|
| Power | 240 vac 50/60 hz, 3 a (240 vac) 6a (110 vac) | |
| Dimensions | 45 x 41x 21 cm | 80 x 62 x 41 cm |
| Weight | 13 kilos | 37 kilos |



VIP会员