Minisim 1000 patient simulator is a multi-parametric patient simulator designed to test and validate the performance of the patient custody system quickly and easily. This microprocessor-based instrument is a soft touch keyboard and visual lcd screen drive menu。
It simulates ecgs, arrhythmia, blood pressure, breathing and temperature, and performance wave shapes that include sine waves, triangle waves, pulse waves and square waves。
- 12 contemperature map
- Forty-five cardiac disorders
- An electrocardiogram performance waveform
- 2 blood pressure simulation
- Respiration/respiration pause
- Temperature
- Easy to use
- Small size
- Menu driver operations
- Large lcd monitor
- Touch keyboard
- Battery or exchange power
Ecg output: 12 conduits, independent output with reference rl。
Normal heart rate (nsr)
Ecg rates: 30, 60, 70, 80, 90, 100, 120, 150, 180, 210, 240, 270, 300 and 350 bpm. 0. 5% accuracy
Range on route ii: 0. 15, 0. 3, 0. 5, 1. 0, 2. 0, 3. 0, 4. 0 and 5. 0 mv. Line 1 is 0. 6 x lead ii, line iii is 0. 4 x lead ii, line v is the same as line ii
High level: 500 x low level output on head ii
Accuracy: 2% (1-5mv)
Performance waveform (perf)
Sine waves, square waves, triangle waves, pulses
Frequency: 0. 1 to 0. 9 by 0. 1 hz
1. 0 hz step 1. 0 to 9. 0
10 hz step 10 to 100
Precision: 1%
Range: 0. 1, 0. 2, 0. 5, 1. 0, 2. 0, 3. 0, lead ii
Pulse voltage of 5. 0 mv: repeat 1 ms amplitude of 20 ms vein at 4 seconds intervals charge
Rwd
Width: 10, 40, 50, 60, 70, 80, 90, 100, 110 and 120 ms. On route two
Range: 0. 15, 0. 3, 0. 5, 1. 0, 2. 0, 3. 0, 5. 0 mv。
Pacer (pcr)
Ap: cardiac pacemaker set to 70 bpm
Asp: interstep pacemaker
Qrs range: 1 mv
Pcr: pacemaker pulse only charge
Trigger width: 0. 1, 0. 2, 0. 5, 1. 0, 1. 5 and 2. 0 ms
Range: 0. 15, 0. 3, 0. 5, 1. 0, 2. 0, 3. 0, 4. 0 and 5. 0 mv
Vp: cardiac pacemaker set to 70 bpm
Avp: cardiac pacemaker set to 70 bpm. Qrs amplitude: 1 mv pacemaker
Amplitude: -2mv
Breathe
Baseline resistance: 250, 500, 750 and 1,000 om
Incremental resistance: 0. 1, 0. 5, 1. 0 and 1. 5 om
Accuracy: 20%
Rates: 15, 30, 60 and 120 bpm
Precision: 1%
Respiration pause: off, continuous, 12 and 32 seconds
Line configuration: lines i and ii
Blood pressure
Resist: 350 om
Incentives: 2 to 16 volts
Sensitivity: 5 mv / v / mmhg
Static
0, 5, 10, 20, 25, 30, 40, 50, 100, 150, 200 and 300 mmhg dynamics: 50/10, 60/20, 70/30, 80/40, 100/60 and 120/80 mm mercury column. Dynamic wavelength tracking all nsr speeds
Temperature
Compatibility: ysi 400 and 700 series。
Temperature: 25° and 37°c (77° and 98. 6°f)
Accuracy: 2% of the set value