A compact, small size rigel uni-therm automatically performs all esu-related performance tests。
Large current testing up to 8 ampere rms
No sensor load from 0 - 5115 times
All iec 60601-2-2 leak tests
Screen power configuration file analysis
Board automation and data capture
Full colour user interface
Match:
Gb9706. 4 (iec60601-2-2) medical electrical equipment
Jjf 1217-2009 high frequency electric knife calibration code
The new high-performance rigel uni-therm electrocutor analyser can accurately measure the performance of the electrosurgery generator, including high frequency leaks, large currents, distribution and patient back warning tests。
The uni-therm electrosurgery analyser meets the iec 60601-2-2 standard, is capable of verifying and calibrating all major electrosurgery generators, and automatically guides users to complete all testing procedures and, most importantly, is safe。
The unique high-power low-power sensor load team was able to achieve performance tests up to 8 ampe rms, with a space ratio of up to 100 per cent, thus making the analyser an extremely common test instrument for regular and high-power electrosurgery generator calibration and performance testing. The load group allows tests to be carried out in a range of 0-5115 times with a resolution of more than 1,000 paces and provides a highly detailed power curve. Rigel uni-therm also promotes safe working practices by providing all necessary resisters in one shell。
Returning electrode monitoring capabilities can be performed manually or automatically. High-precision bitmeters can be tested within one resolution range and are ideal when alert settings are based on resistance to change tests。
The rigel uni-therm sword analyser product functions include plate memory, testing automation, integrated data management facilities and various built-in resistors, while large-scale full colour screens display easily followed step descriptions to ensure that the correct connection to the equipment is being tested。
Bluetooth technology allows wireless connections to personal computers and other equipment to quickly and easily download test data, upload power curves of electrosurgery equipment and test programs for manufacturers. The output wave shape can be checked through the built-in oscilloscope output and can easily confirm the required wave shape。
The rigel uni-therm escalator is part of a comprehensive high performance specialized biomedical test equipment provided by rigel medical。
Rigel uni-therm high-electric current surgical analyser
| Specification of the product of the rigel uni-therm escalator | |
|---|---|
| Power measurement | Apply wave value |
| Rating power | 0 - 500w (rms) |
| Precision | ± (1w + 5% value) |
| Percentage | 100% to 60 seconds |
| Load bank | 0 - 5115 times |
| Resolution | Five |
| Electricity (peak) | 0 - 10kv (peak) - closed load only |
| Precision | ± (10% of reading + 15v) Only measurements between active and dispersed electrodes of closed loads。 |
| Voltage | 0 - 700v (rms) |
| Precision | (5v + 10%) |
| Current | 0 - 6000ma (rms) |
| Precision | ± (2% reading + 10ma) |
| Championship factor | 1. 4 - 20 (vpeak / v rms) One of the two peak voltage measurements is used to calculate |
| Rms bandwidth | |
|---|---|
| Only instruments | 30 hz to 10 mhz (-3 db) |
| Load: | 30 hz to 2. 5 mhz (-3 db) |
| Variable load | 0 - 5115, step five |
| Precision | 1% of r |
| Load arrays | Ceramic electrical resistance (non-electrical) |
| Measuring delay | Step switch delay can be selected between 200 and 5,000 ms (10 ms resolution) |
| Radio frequency leak (hf leak) | |
|---|---|
| Active | From the positive part to the earth |
| Passive | From the plate - the container - to the earth |
| Load | Variables see power measurements fixed 2 x 200 times |
| Return to electrode monitoring | |
|---|---|
| Scope | 1,475 times, step @1 step into electric power drive |
| Alarm clock repository | High and low, manual confirmation |
| Ranging | Manual or automatic |
| Output connector | |
|---|---|
| Remote footstep switch control (cut) | 2 x 4mm - yellow, single relay contact |
| Remotely controlled footstep switches (coag) | 2 x 4mm - blue, single relay contact |
| Hf leakage and power measurement | Through the 4 mm plug |
| Usb | Pc download/future pc software |
| Oscilloscope output | Minimum input for 5v / a, 100ma radio frequency current |
| Segregation |
|---|
| 10kv separation between measuring equipment and shell |
| Low frequency filter |
|---|
| 100 hz filter to avoid low frequency interference or interference |
| Storage and recall | |
|---|---|
| Memory | About 5,000 records |
| Production | Csv and sss formats |
| General specifications | |
|---|---|
| Outer size | 370x300x204mm |
| Weight | Ten kilos |
| Working temperature | 15°c to 40°c |
| Storage temperature | 0°c to 50°c |
| Main power | 120/230 vac + 10%;48 to 66 hz, 35va |
| Spare parts number | 398a910 |
| Services and warranties |
|---|
| Uni-therm has 24 months of warranty for free upgrades with terms and conditions。 A series of med-ecare schemes is also available。 |