Calibermri 208 prostate modelling module, calibermri 208 prostate mri simulator, consisting of:
- T1 simulator*
- T2 simulator*
- Adc simulation*
- The patented 10 element lcd (lc) thermometer, mr readable, for temperature compensation of 15°c to 24°c。
- A resolution blade with holes, ranging from 0. 44 mm to 1. 2 mm in diameter。
Mri standardization has become easier through the cmri patent*lc mr readable thermometers developed in collaboration with nist. Qmri improvements can now be standardized between 15°c and 24°c while scanning proliferation models。

Calibermri 208 prostate model, calibermri 208 prostate mri model specification:
- Qmri specialists consist of seven cylinders filled with pvp water solutions ranging from 0% to 50% w/w, which allow accurate and precise adc measurements。
- Nist traceable t1 solution, measured under 3t, occupies 7 cylinders, covering the 1989 ms - 125. 9 ms t1 range。
- Nist traceable t2 solution occupied 7 cylinders covering the t2 range of 278. 1 ms – 32. 0 ms
Calibermri 208 prostate model, calibermri 208 prostate mri characteristics for assessment
- B1 and b0 evenness
- Geometry
- Gradient range
- System centre frequency drift (short duration)
- Resolution (high contrast detectability and snr (low contrast detectability)
- T1 and t2 measurements are accurate and accurate
- Accuracy and precision of the watch proliferation factor (adc)

Development in partnership with nist and minnesota university。
190 mm x 280 mm long (axis) cylinder shell with 0. 4% saline water solution。
Rotable prostate simulation modules include:
- T1 emulator*: 7 vials with nist traceable nicl, solution from ~120 ms to ~1900 ms。
- T2 simulator*: 7 vials containing nist traceable mncl solutions ranging from ~30 ms to ~270 ms。
- Adc simulation*: 7 vials with pvp solutions measured by nist, adc0 values from ~0. 3x10-° mm'sec to ~2x10-° mm? Sec。
- Patented lcd (lc) thermometer, mr readable, temperature compensation range from 15°c to 24°c。
- Resolution inserts the diameter of the hole pattern from 00. 4 mm to 1. 2 mm。
Optional tube for inserting 40 mm and 25 mm radio frequency recternal (erc)
Compatible with qcal-mr qa/oc software。