Image collection and image processing at multiple levels of capability are improving the visualization of computer fault photography。
Qrm-sct spectra ct module, qrm-sct ct test module, can be used to test different types of ct machine through double energy, multi-energy or photometer settings。
The 100 mm cylinder consists of 4 20 mm diameter holes, which are used to test rods for different solid tissue equivalents or can be filled with water or film agents。
Modular models include a different set of solid rods that are equivalent to soft tissues such as iodine, calcium-hydroxyphosphate, water and fat, muscle, bones and lungs. A mass density, electron density and a valid atomic sequence (z) are given for each material。
For testing multi-energy or photometer ct protocols and reprocessing techniques
D-100 compatible
Modular model for multi-energy or photometric settings for qa and research
It's an ideal option for material representation and organization of an equivalent amount of material
To provide different pathologies for spectrum ct
Ca ++ and iodine rods
Spectrum ct-modules can be used to test different types of ct-modules through dual energy, multi-energy or photometer settings。
The 100 mm cylinder slot consists of four holes with a diameter of 20 mm, which are used to contain test rods for different solid tissue equivalents and fillable rods that can be used with water or a reflection agent。
Modular models include a different set of solid rods that are equivalent to soft tissues such as iodine, calcium-hydroxyphosphate, water and fat, muscle, bones and lungs. Provides mass density, electronic density and effective atomic sequences for each material。
Diameter: 100 mm
Length: 103 mm
Weight: 1. 0 kg
Bar diameter: 20 mm
Length of pole: 103 mm
Water equivalent cylinder slots with 4 holes (calculated)
4 plugin ctwater (0 hu@70-150 kv)
12 test sticks:
Fatty tissue (icru 44), muscle tissue (icru 44), lung tissue (icru 44), liver tissue (icru 44)
4 different iodine rods 2, 5, 10 and 15 mg i/cm3
4 different ca-ha sticks 100, 200, 400 and 800 mg ha/cm3
2 fillable sticks/pipes
The accompanying d160 mm and d320 mm water emulator circles are optional, the filling rods are optional, the icru 44 material tissue plugins are optional, and bone tissue materials are optional select
Qrm-sct spectrum ct module, qrm-sct ct test module characteristics:
1, test the accuracy and consistency of spectral ctTwo, test scanner performance
3. Material display and quantification for organizational equivalents
4. Decomposition of iodine and calcium
Five, test spectral ct reprocessing