Two inner cavity scans with an inner diameter of 3 mm in the center of the panel. Chatoons, which are used to insert inner cavity probes for ultrasonic scans, are specially designed to detect visual characterization parameters of the b hyperimage device attached to the cavity, including blind areas, side-to-side (horizontal) resolution, axis (vertical) resolution, side-to-side geometry accuracy, axis-to-geometry accuracy and diameter error in the simulation stove. The side panel on the right and left sides of the model has a probe tunnel with an inner diameter of 3 mm for the measurement of the geometrical position of the retreat direction resolution and retreat direction. The physical appearance and internal structure are illustrated in figure 1。
1. The four walls and bottoms are assembled in organic glass processing and plastic sheeting with instructions and decorations on the outer surface of the four walls
4. The module contains tm material meeting national standards as a standard acoustic medium。
5. A cavity probe with a diameter of 3 mm in the centre of the acoustic model, 60 mm and 120 mm above the acoustic window, which is a penetrator channel capable of ultrasounding through the hole. The side panel on the right and left sides of the model has a 3 mm internal diameter for the measurement of the geometrical position of the retreat direction and direction. The channel is the channel through the left and right panels, and the inner cavities can be scanned from the right and right sides。
(1) o1-o3 probe
The surface plate of the model is open with two probe holes with a diameter of 3 mm, located in the centre of the first half of the panel and the centre of the second half. Proof o1 is surrounded by axle-side resolution target, side-to-side target and vertical target. Zachon o2 is surrounded by blind target sets, geometrically deformed target groups, cystic cookstoves, tumours, cysts and imitates. Proof o3 is located on both sides of the side panel to the left and right of the body model and is used to measure the accuracy of the geometric position in the direction of retreat and in the direction of retreat. The channel is the channel through the left and right panels, and the inner cavities can be scanned from the right and right sides。
(2)a: blind targets
The blind target group is set around chacon o2, with a total of 8 target lines set every 45 degrees, starting at 1 mm above the outer edge of chacon o2 with a 1 mm velocity, respectively(3) b1-b4: axes side resolution target group
The axle-side resolution target group is placed around zachon o1, with a circle of 5,10,20,30 mm at the edge of zachon o1 from the branch, respectively, i. E. B1 and b2, b3 and b4 four groups, each of which is spread around a radius of 6. 5 mm, 11. 5 mm, 21. 5 mm and 31. 5 mm from the centre of the zachon. The range (long lines between the two target lines) is 3,2,1,0. 5 mm in descending order。The target group has a radius of two adjacent target lines in its axis and branch, with a radius of 0. 5,1,2,3 mm at the centre of the adjacent target line at the centre of the centre of the centre of the centre. For example, the directional resolution of the axis at a radius of 6. 5 mm at the centre of the probe, or 5 mm on the edge of the probe, is shown in figure 3. The intervals between target lines in the axis resolution target range are the direct range of the target line and the length of target lines in the side resolution target group (non-circle arc long)。
(4) c1 - c3: side-to-side target group (for the detection of side-to-side (horizontal) geometric accuracy)
(5) d1-d2: vertical targets
Longitudinal target group d1 is positioned around zachon o1 and is located on 6 of the target-bearing line at the margin of 5, 10, 20, 30, 40 and 50 mm to the left of zachon o1。Longitudinal target group d2 is placed around zachon o1, with a total of 4 target lines located at 5, 10, 20, 30 mm below the positive edge of zachon o1。
(6) j: image geometric alteration target group
The image geometric aberration target group is set around zachon o2 and is located on four of the target-bearing lines, at the top and bottom of the square at 10 mm from the edge of zachon o2。(7)h1 - h4: resolution target in the direction of the retreat of the scout hole group
The retreat direction resolution, in order to detect the energy exchanger's retreat in the direction of the probe cone axis, shows the minimum distance between the two target lines of the two clear echo signals. Rescission of the resolution target set in the vicinity of chamac o3. The h1-h4 target group is located on the edge of os3, 5, 10, 15 and 20 mm from the centre. The resolution self-discretionary detection hole of each group of targets is 3,2,0. 5 mm from the inner direction。(8)h5-h6: sound window retreat direction resolution target group
(9) i1-i2: geometric target in the direction of the retreat of the scout hole group
The geometrical position in the direction of the retreat was positioned in the vicinity of cham-o3. The i1 target group is located at 5 and 10 mm from the edge of the o3 tunnel at the centre and at 15 and 20 mm from the edge of the o3 tunnel at the centre. The range of the target line per group is 5 mm。(10) i3: geometric target in the direction of sound window retreat group
The accuracy of the target group i3 in the direction of the sound window is set at 5 mm, 15 mm, 25 mm depth under the sound window. The target range of the i3 target group is 5 mm。7. Simulation of disease stoves
(1) e1-e3: replicates
The fungus e1-e3 are placed around zachon o2 and are embedded in three cylindrical simulation stoves, all cylindrical, with a diameter of 2,4,6 mm, all of which are parallel to the target line, with a axis located at 5,10,30 mm above the centre of the scan。