Assess the full range of clinical imaging frequency
The cirs model ats 539 multi-purpose phantom is a simple, comprehensive method for assessing imaging systems within a full range of clinical imaging frequencies (2 mhz to 18 mhz)。
Phantom designs a combination of organizational simulation target structures for single-wire objectives and different sizes and contrasts for distance measurements. The cylindrical sample target structure is vertically organized to allow the entire range to be displayed in one view. Due to the acoustic similarity of the background material and target structure, hypocrisy caused by distortion, shadow or enhancement was eliminated. Provides six greyscale targets ranging from +15 to 15 db to assess the dynamic scope of the system and the greyscale treatment performance。
Cirs ats 539 multipurpose model, ats 539 ultrasound model, ats 539 b superperformance model assessment indicator:
Even
Depth of penetration
Beam contour/focus area/horizontal response width
Vertical distance measurement
Horizontal distance measurements
Axis and horizontal resolution
High range resolution
Contrast resolution
Grey contrast sensitivity
Blind zone assessment
Cirs ats 539 multipurpose model, ats 539 ultrasound model, ats 539 b superperformance model product specification:
Dimensions: 23. 4 cm x 20. 5 x 9. 5 cm (9" x 8" x 4)
Weight: 7 lbs (3. 1 kg)
Shell material: pvc
Wall thickness: 1 cm
Windows dimensions: (2) 15. 5 x 7. 5 cm; (2) 14. 0 x 7. 5 cm
Organisation simulation material: polyurethane rubber
Polyurethane properties:
Condensation point: <-40°c
Melting point: 100°c or more
Sound speed: 1450 m/s ± 1. 0% 23°
Decline: 0. 5 db/cm/mhz±10%
Line target: material: monosnilon diameter: 0. 12 mm
Vertical distance group (17 targets):
Depth range 1 - 18 cm, 1 cm
Horizontal distance group: linear probes (5 targets):
Five centimeters from the surface of the scanned window, two centimeters away
Horizontal distance group: sector scan probes (5 targets):
Depth 13 cm from the scanned surface of the acoustic window, 2 cm
Dead areas (9 targets):
Depth range 20 - 100 mm, 1 cm
Axial and horizontal resolution groups (11 objectives):
Horizontally 1 mm
Distances 4, 3, 2, 1, 0. 5 and 0. 25 mm
Depth range 7, 11, 4, 16 cm
Photographic target (elastema) (group 5):
Type: no echo, cylinder
Clusters 2, 3 and 4 (17 objectives)
Dimensions: 2, 3, 4 mm
Depth: 1 - 17 cm
Distance: 1cm
Clusters 6 and 8 (8 objectives)
Dimensions: 6,8 mm
Depth: 1 - 17 cm
Distance: 2 cm
Grayscale target (6 targets):
Type: echo, cylinder
Spacing: 2 cm (centre to centre)
Depth comparison: 4 cm contrast: +15, +6, +3, 6-15 db relative to background